differential scanning calorimetry

DSC 204 F1 Phoenix

The premium DSC for all requirements 

Muestra el equipo actual

The versatile Premium Differential Scanning Calorimeter, DSC 204 F1 Phoenix, fascinates with its unique concept: all essential operational components are integrated into the instrument and optional add-ons can be installed anytime on site.

The DSC measuring cell consists of a cylindrical high-conductivity silver block with an embedded heating coil for broad thermal symmetry (3D symmetry) in the sample chamber, the cooling ports for liquid nitrogen or compressed air cooling and a cooling ring for connection of the intracooler (also with simultaneous liquid nitrogen cooling). The gas-tight construction and integrated mass flow controllers for the purge and protective gases allow coupling to an FT-IR or MS for gas analysis.

Tau-sensor
µ-sensor

Exchangeable sensors are tailored to their tasks:

With its disk-shaped silver carrier plate and supersensitive thermosensors of nickel-chromium constantan, the Tau-sensor offers a high level of calorimetric sensitivity along with extremely short signal time constants of only 0.6 seconds, which guarantees good separation of overlapping thermal effects.

The µ-sensor stands out for its high level of calorimetric sensitivity, never before achieved with a Differential Scanning Calorimeter, DSC. For example, it is extremely well suited for pharmaceutical applications with small sample weights.

Technical Data

Temperature range
-180°C to 700°C
Heating rates
0.001 K/min to 200 K/min
Cooling rates
max. 200 K/min

Compressed air cooling:
RT to 700°C

Intracooler for the extended range:
-85°C to 600°C

Automatically controlled LN2 cooling:
-180°C to 700°C

Indium Response Ratio (*):
>160 mW/K
(*) Peak Height [mW]/Peak with at 50% Height [K] of Indium melting peak

Gas control for 2 sample gases and 1 protective gas through integrated mass flow controller and software

Automatic sample changer:
for 192 samples and reference crucibles (optional)